Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/25216
Title: Motor neuroprosthesis implanted with neurointerventional surgery improves capacity for activities of daily living tasks in severe paralysis: first in-human experience.
Austin Authors: Oxley, Thomas J;Yoo, Peter E;Rind, Gil S;Ronayne, Stephen M;Lee, C M Sarah;Bird, Christin;Hampshire, Victoria;Sharma, Rahul P;Morokoff, Andrew;Williams, Daryl L;MacIsaac, Christopher;Howard, Mark E ;Irving, Lou;Vrljic, Ivan;Williams, Cameron;John, Sam E;Weissenborn, Frank;Dazenko, Madeleine;Balabanski, Anna H;Friedenberg, David;Burkitt, Anthony N;Wong, Yan T;Drummond, Katharine J;Desmond, Patricia;Weber, Douglas;Denison, Timothy;Hochberg, Leigh R;Mathers, Susan;O'Brien, Terence J;May, Clive N;Mocco, J;Grayden, David B;Campbell, Bruce C V;Mitchell, Peter;Opie, Nicholas L
Affiliation: Synchron, Inc, Campbell, California, USA
Interventional Cardiology, Cardiovascular Medicine Faculty, Stanford University, Stanford, California, USA
Battelle Memorial Institute, Columbus, Ohio, USA
Department of Mechanical Engineering and Neuroscience Institute, Carnegie Mellon University, Pittsburgh, Pennsylvania, USA
Center for Neurotechnology and Neurorecovery, Department of Neurology, Massachusetts General Hospital, Harvard Medical School, Harvard University, Cambridge, Massachusetts, USA
Department of Electrical and Computer Systems Engineering, Monash University, Clayton, Victoria, Australia
Neurosurgery, The Mount Sinai Health System, New York, New York, USA
Vascular Bionics Laboratory, Departments of Medicine, Neurology and Surgery, Melbourne Brain Centre at the Royal Melbourne Hospital, The University of Melbourne, Melbourne, Victoria, Australia
Medicine, University of Melbourne, Parkville, Victoria, Australia
Neurology, Royal Melbourne Hospital, Melbourne, Victoria, Australia
Florey Institute of Neuroscience and Mental Health, University of Melbourne, Parkville, Victoria, Australia
Neurology, Melbourne Health, Parkville, Victoria, Australia
Synchron, Inc, Campbell, California, USA
Institute of Biomedical Engineering, Oxford University, Oxford, Oxfordshire, UK
Neurology, Calvary Health Care Bethlehem, South Caulfield, Victoria, Australia
Neurosurgery, Melbourne Health, Parkville, Victoria, Australia
Vascular Bionics Laboratory, Departments of Medicine, Neurology and Surgery, Melbourne Brain Centre at the Royal Melbourne Hospital, The University of Melbourne, Melbourne, Victoria, Australia
Department of Biomedical Engineering, University of Melbourne, Melbourne, Victoria, Australia
Radiology, Melbourne Health, Parkville, Victoria, Australia
Respiratory Medicine, Melbourne Health, Parkville, Victoria, Australia
Institute for Breathing and Sleep
Intensive Care Unit, Melbourne Health, Parkville, Victoria, Australia
Anaesthesia, Melbourne Health, Parkville, Victoria, Australia
Issue Date: Feb-2021
Date: 2020-10-28
Publication information: Journal of neurointerventional surgery 2021; 13(2): 102-108
Abstract: Implantable brain-computer interfaces (BCIs), functioning as motor neuroprostheses, have the potential to restore voluntary motor impulses to control digital devices and improve functional independence in patients with severe paralysis due to brain, spinal cord, peripheral nerve or muscle dysfunction. However, reports to date have had limited clinical translation. Two participants with amyotrophic lateral sclerosis (ALS) underwent implant in a single-arm, open-label, prospective, early feasibility study. Using a minimally invasive neurointervention procedure, a novel endovascular Stentrode BCI was implanted in the superior sagittal sinus adjacent to primary motor cortex. The participants undertook machine-learning-assisted training to use wirelessly transmitted electrocorticography signal associated with attempted movements to control multiple mouse-click actions, including zoom and left-click. Used in combination with an eye-tracker for cursor navigation, participants achieved Windows 10 operating system control to conduct instrumental activities of daily living (IADL) tasks. Unsupervised home use commenced from day 86 onwards for participant 1, and day 71 for participant 2. Participant 1 achieved a typing task average click selection accuracy of 92.63% (100.00%, 87.50%-100.00%) (trial mean (median, Q1-Q3)) at a rate of 13.81 (13.44, 10.96-16.09) correct characters per minute (CCPM) with predictive text disabled. Participant 2 achieved an average click selection accuracy of 93.18% (100.00%, 88.19%-100.00%) at 20.10 (17.73, 12.27-26.50) CCPM. Completion of IADL tasks including text messaging, online shopping and managing finances independently was demonstrated in both participants. We describe the first-in-human experience of a minimally invasive, fully implanted, wireless, ambulatory motor neuroprosthesis using an endovascular stent-electrode array to transmit electrocorticography signals from the motor cortex for multiple command control of digital devices in two participants with flaccid upper limb paralysis.
URI: https://ahro.austin.org.au/austinjspui/handle/1/25216
DOI: 10.1136/neurintsurg-2020-016862
ORCID: 0000-0001-7772-1496
Journal: Journal of neurointerventional surgery
PubMed URL: 33115813
Type: Journal Article
Subjects: brain
device
intervention
technology
vein
Appears in Collections:Journal articles

Show full item record

Page view(s)

70
checked on Dec 24, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.