Title: A meta-analysis of the accuracy and precision of the ultrasonic cardiac output monitor (USCOM).
Austin Authors: Chong, S W;Peyton, Philip J 
Affiliation: Department of Anaesthesia, Austin Hospital, Melbourne, Australia
Issue Date: 29-Aug-2012
Publication information: Anaesthesia 2012; 67(11): 1266-71
Abstract: The ultrasonic cardiac output monitor is a non-invasive, continuous-wave Doppler monitor that can be used to measure cardiac output. Its accuracy and precision in peri-operative and critical care has been assessed with varied results. We conducted a meta-analysis to calculate a pooled mean bias, precision and percentage error for this device. Six studies were suitable for inclusion into a pooled parametric analysis consisting of 320 measurements. The mean weighted bias was -0.39 l.min(-1) (95% CI -0.25 to -0.53 l.min(-1) ), precision 1.27 l.min(-1) and percentage error 42.7% (95% CI 38.5-46.9%). The ultrasonic cardiac output monitor achieved similar agreement with bolus thermodilution to that of other minimally invasive methods of peri-operative cardiac output monitoring, and may have a useful role in cardiac output monitoring.
Gov't Doc #: 22928650
URI: https://ahro.austin.org.au/austinjspui/handle/1/11552
DOI: 10.1111/j.1365-2044.2012.07311.x
Journal: Anaesthesia
URL: https://pubmed.ncbi.nlm.nih.gov/22928650
Type: Journal Article
Subjects: Cardiac Output
Echocardiography, Doppler
Humans
Monitoring, Intraoperative.methods
Reproducibility of Results
Thermodilution