Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/11552
Title: A meta-analysis of the accuracy and precision of the ultrasonic cardiac output monitor (USCOM).
Austin Authors: Chong, S W;Peyton, Philip J 
Affiliation: Department of Anaesthesia, Austin Hospital, Melbourne, Australia
Issue Date: 29-Aug-2012
Publication information: Anaesthesia 2012; 67(11): 1266-71
Abstract: The ultrasonic cardiac output monitor is a non-invasive, continuous-wave Doppler monitor that can be used to measure cardiac output. Its accuracy and precision in peri-operative and critical care has been assessed with varied results. We conducted a meta-analysis to calculate a pooled mean bias, precision and percentage error for this device. Six studies were suitable for inclusion into a pooled parametric analysis consisting of 320 measurements. The mean weighted bias was -0.39 l.min(-1) (95% CI -0.25 to -0.53 l.min(-1) ), precision 1.27 l.min(-1) and percentage error 42.7% (95% CI 38.5-46.9%). The ultrasonic cardiac output monitor achieved similar agreement with bolus thermodilution to that of other minimally invasive methods of peri-operative cardiac output monitoring, and may have a useful role in cardiac output monitoring.
Gov't Doc #: 22928650
URI: https://ahro.austin.org.au/austinjspui/handle/1/11552
DOI: 10.1111/j.1365-2044.2012.07311.x
Journal: Anaesthesia
URL: https://pubmed.ncbi.nlm.nih.gov/22928650
Type: Journal Article
Subjects: Cardiac Output
Echocardiography, Doppler
Humans
Monitoring, Intraoperative.methods
Reproducibility of Results
Thermodilution
Appears in Collections:Journal articles

Show full item record

Page view(s)

22
checked on Oct 7, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.