Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/32559
Title: Ventricular fibrillation episodes related to managed ventricular pacing algorithm in a patient with mitral valve prolapse - "Arrhythmic MVP due to MVP".
Austin Authors: Chen, Janet X C;Fong, Jenny;Lim, Han S ;Teh, Andrew W ;Han, Hui-Chen 
Affiliation: Austin Health, Victoria, Australia.
Eastern Health Clinical School, Monash University, Victoria, Australia
University of Melbourne Clinical School
Monash Medical Center, Victoria, Australia
Issue Date: 13-Jun-2023
Date: 2023-03
Publication information: Journal of Electrocardiology 2023; 79
Abstract: We report a case of an implant cardiac defibrillator (ICD) patient who an ICD implanted for ventricular fibrillation (VF) related to mitral valve prolapse. He has 2 episodes of VF in his device lives. First episode of VF in year 2016 was initiated after a pause related to the MVP™ algorithm with a critically timed PVC. MVP™ was turned off which prevented further VF episodes. However, MVP™ was turned back on at the device replacement in 2018. A second VF episode developed with similar mechanism in 2021 and MVP™ was subsequently turned off with no further VF recorded. This case highlights the importance of recognizing the mechanism of initiation of tachy-arrhythmia episodes and serves as an important reminder regarding optimization of device settings at the time of replacement.
URI: https://ahro.austin.org.au/austinjspui/handle/1/32559
DOI: 10.1016/j.jelectrocard.2023.03.008
ORCID: 
Journal: Journal of Electrocardiology
Start page: 58
End page: 60
PubMed URL: 36948089
ISSN: 1532-8430
Type: Journal Article
Subjects: Device settings
ICD
Managed ventricular pacing
Mitral valve prolapse
Ventricular fibrillation
Appears in Collections:Journal articles

Show full item record

Page view(s)

40
checked on Jul 3, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.