Title: | Autocycling During Noninvasive Positive Pressure Ventilation Producing a Prolonged Severe Apnea and Syncope. | Austin Authors: | Mu, Susana;Rautela, Linda ;Howard, Mark E ;Hannan, Liam M | Affiliation: | University of Melbourne, Parkville, Victoria, Australia Institute for Breathing and Sleep Northern Health, Epping, Victoria, Australia Austin Health Victorian Respiratory Support Service |
Issue Date: | Apr-2019 | Date: | 2019-04 | Publication information: | Journal of Clinical Sleep Medicine 2019; 15(4): 663-665 | Abstract: | Autocycling is a form of patient-ventilator asynchrony that can occur during mechanical ventilation. In the case described, autocycling during noninvasive ventilation led to severe hyperventilation and subsequently produced a prolonged central apnea that resulted in syncope. This case represents the first description of a severe adverse event from autocycling during noninvasive ventilation. | URI: | https://ahro.austin.org.au/austinjspui/handle/1/24902 | DOI: | 10.5664/jcsm.7738 | Journal: | Journal of Clinical Sleep Medicine | PubMed URL: | 30952221 | Type: | Journal Article | Subjects: | autocycling noninvasive positive pressure ventilation patient-ventilator asynchrony syncope |