Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/17684
Full metadata record
DC FieldValueLanguage
dc.contributor.authorEbmeier, S J-
dc.contributor.authorBarker, M-
dc.contributor.authorBacon, M-
dc.contributor.authorBeasley, R C-
dc.contributor.authorBellomo, Rinaldo-
dc.contributor.authorKnee Chong, C-
dc.contributor.authorEastwood, Glenn M-
dc.contributor.authorGilchrist, J-
dc.contributor.authorKagaya, H-
dc.contributor.authorPilcher, J-
dc.contributor.authorReddy, S K-
dc.contributor.authorRidgeon, E-
dc.contributor.authorSarma, N-
dc.contributor.authorSpragas, S-
dc.contributor.authorTanaka, Aiko-
dc.contributor.authorTweedie, M-
dc.contributor.authorWeatherall, M-
dc.contributor.authorYoung, P J-
dc.date.accessioned2018-05-08T23:56:50Z-
dc.date.available2018-05-08T23:56:50Z-
dc.date.issued2018-05-
dc.identifier.citationAnaesthesia and Intensive Care 2018; 46(3): 297-303-
dc.identifier.issn0310-057X-
dc.identifier.urihttps://ahro.austin.org.au/austinjspui/handle/1/17684-
dc.description.abstractThe influence of variables that might affect the accuracy of pulse oximetry (SpO<sub>2</sub>) recordings in critically ill patients is not well established. We sought to describe the relationship between paired SpO<sub>2</sub>/SaO<sub>2</sub> (oxygen saturation via arterial blood gas analysis) in adult intensive care unit (ICU) patients and to describe the diagnostic performance of SpO<sub>2</sub> in detecting low SaO<sub>2</sub> and PaO<sub>2</sub>. A paired SpO<sub>2</sub>/SaO<sub>2</sub> measurement was obtained from 404 adults in ICU. Measurements were used to calculate bias, precision, and limits of agreement. Associations between bias and variables including vasopressor and inotrope use, capillary refill time, hand temperature, pulse pressure, body temperature, oximeter model, and skin colour were estimated. There was no overall statistically significant bias in paired SpO<sub>2</sub>/SaO<sub>2</sub> measurements; observed limits of agreement were +/-4.4%. However, body temperature, oximeter model, and skin colour, were statistically significantly associated with the degree of bias. SpO<sub>2</sub> <89% had a sensitivity of 3/7 (42.9%; 95% confidence intervals, CI, 9.9% to 81.6%) and a specificity of 344/384 (89.6%; 95% CI 86.1% to 92.5%) for detecting SaO<sub>2</sub> <89%. The absence of statistically significant bias in paired SpO<sub>2</sub>/SaO<sub>2</sub> in adult ICU patients provides support for the use of pulse oximetry to titrate oxygen therapy. However, SpO<sub>2</sub> recordings alone should be used cautiously when SaO<sub>2</sub> recordings of 4.4% higher or lower than the observed SpO<sub>2</sub> would be of concern. A range of variables relevant to the critically ill had little or no effect on bias.-
dc.language.isoeng-
dc.subjectintensive care medicine-
dc.subjectmonitoring-
dc.subjectpulse oximetry-
dc.subjectpulse oximetry, monitoring: oxygen saturation tones-
dc.titleA two centre observational study of simultaneous pulse oximetry and arterial oxygen saturation recordings in intensive care unit patients.-
dc.typeJournal Article-
dc.identifier.journaltitleAnaesthesia and Intensive Care-
dc.identifier.affiliationIntensive Care Unit, Austin Health, Heidelberg, Victoria, Australia-
dc.identifier.affiliationIntensive Care Research, Medical Research Institute of New Zealand;, Wellington, New Zealand-
dc.identifier.affiliationIntensive Care Unit, Wellington Regional Hospital, Wellington, New Zealand-
dc.identifier.orcid0000-0002-1650-8939-
dc.identifier.pubmedid29716488-
dc.type.austinJournal Article-
local.name.researcherBellomo, Rinaldo
item.grantfulltextnone-
item.openairetypeJournal Article-
item.languageiso639-1en-
item.fulltextNo Fulltext-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.cerifentitytypePublications-
crisitem.author.deptIntensive Care-
crisitem.author.deptData Analytics Research and Evaluation (DARE) Centre-
crisitem.author.deptIntensive Care-
Appears in Collections:Journal articles
Show simple item record

Page view(s)

42
checked on Nov 19, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.