Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/13357
Full metadata record
DC FieldValueLanguage
dc.contributor.authorKalman, J Men
dc.contributor.authorPower, JMen
dc.contributor.authorChen, J Men
dc.contributor.authorFarish, S Jen
dc.contributor.authorTonkin, Andrew Men
dc.date.accessioned2015-05-16T03:11:37Z
dc.date.available2015-05-16T03:11:37Z
dc.date.issued1993-10-01en
dc.identifier.citationJournal of the American College of Cardiology; 22(4): 1199-206en
dc.identifier.govdoc8409061en
dc.identifier.otherPUBMEDen
dc.identifier.urihttps://ahro.austin.org.au/austinjspui/handle/1/13357en
dc.description.abstractWe assessed the feasibility of low energy endocardial defibrillation in a canine model of atrial fibrillation, comparing catheters with large surface area electrodes and standard electrode catheters, and evaluated the effects of lead configuration and circuit impedance on defibrillation energy requirements.Although recent animal studies have demonstrated the feasibility of low energy endocardial atrial defibrillation, their results have been conflicting with regard to important methodologic aspects.In 14 anesthetized greyhounds, atrial fibrillation was induced by rapid atrial pacing and maintained by vagal stimulation. Two large surface area braided electrode catheters and two standard electrode catheters were introduced percutaneously, one of each, in the right atrial appendage and right ventricular apex. A cutaneous patch electrode was placed on the left thorax. Biphasic shocks synchronized to the ventricular electrogram were used to terminate atrial fibrillation. Seven configurations were evaluated. Three used standard electrodes: proximal atrial cathode to distal atrial, ventricular or cutaneous anode. Four used braided electrodes: three with atrial cathode to ventricular, cutaneous or combined anode; one with ventricular cathode to atrial anode.Defibrillation with standard electrode catheters was associated with high impedance (576 +/- 112 omega) and low success rates for all configurations (28% success at < or = 40 J, no successes at 10 J). Low energy defibrillation was readily achieved with the braided electrodes with significantly lower impedance (75 +/- 13 omega, p < 0.0001). Ventricular fibrillation did not occur. The success rate of cardioversion increased in a dose-response manner, allowing fitting of a sigmoid curve and calculation of energy associated with 50% (ED50) and 90% (ED90) success. The most successful configuration was ventricular cathode/atrial anode (ED50 1.5 +/- 0.4 J), and the least successful was atrial anode/cutaneous patch (ED50 6.5 +/- 3.2 J, p = 0.0001).Low energy atrial defibrillation is feasible using large surface area electrodes but not with standard electrode catheters owing to high impedance. An intracardiac anode provides lower impedance and higher success rates than are provided by a cutaneous anode.en
dc.language.isoenen
dc.subject.otherAnalysis of Varianceen
dc.subject.otherAnimalsen
dc.subject.otherAtrial Fibrillation.pathology.therapyen
dc.subject.otherDisease Models, Animalen
dc.subject.otherDogsen
dc.subject.otherElectric Countershock.instrumentationen
dc.subject.otherElectric Impedanceen
dc.subject.otherElectrocardiographyen
dc.subject.otherElectrodes, Implanted.standardsen
dc.subject.otherEndocardiumen
dc.subject.otherEquipment Designen
dc.subject.otherEvaluation Studies as Topicen
dc.subject.otherFeasibility Studiesen
dc.subject.otherLikelihood Functionsen
dc.subject.otherLogistic Modelsen
dc.titleImportance of electrode design, lead configuration and impedance for successful low energy transcatheter atrial defibrillation in dogs.en
dc.typeJournal Articleen
dc.identifier.journaltitleJournal of the American College of Cardiologyen
dc.identifier.affiliationDepartment of Cardiology, Austin Hospital, University of Melbourne, Parkville, Victoria, Australiaen
dc.description.pages1199-206en
dc.relation.urlhttps://pubmed.ncbi.nlm.nih.gov/8409061en
dc.type.austinJournal Articleen
item.grantfulltextnone-
item.languageiso639-1en-
item.openairetypeJournal Article-
item.cerifentitytypePublications-
item.fulltextNo Fulltext-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
Appears in Collections:Journal articles
Show simple item record

Page view(s)

56
checked on Apr 5, 2025

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.