Please use this identifier to cite or link to this item: https://ahro.austin.org.au/austinjspui/handle/1/10130
Full metadata record
DC FieldValueLanguage
dc.contributor.authorOng, Ten
dc.contributor.authorGillies, M Aen
dc.contributor.authorBellomo, Rinaldoen
dc.date.accessioned2015-05-15T23:29:01Z
dc.date.available2015-05-15T23:29:01Z
dc.date.issued2004-06-01en
dc.identifier.citationCritical Care and Resuscitation; 6(2): 99-101en
dc.identifier.govdoc16566694en
dc.identifier.otherPUBMEDen
dc.identifier.urihttps://ahro.austin.org.au/austinjspui/handle/1/10130en
dc.description.abstractContinuous cardiac output measurement using pulse contour analysis is a technique gaining widespread acceptance in intensive care units. We report a case where a pulse contour analysis computer (PiCCO, Pulsion Medical Systems, Munich, Germany) failed to calibrate in a patient who was undergoing induced hypothermia for anoxic brain injury. Despite several attempts to calibrate, using increased cold injectate volumes and exchanging both the PiCCO device and the arterial catheter, we were unable to correct the calibration problem and hence were unable to monitor cardiac output. Subsequent rewarming of the patient allowed calibration of the arterial waveform and continuous cardiac output measurement. We were unable to find any previous reports of this problem using a PiCCO device, although similar problems with thermodilution cardiac output estimation using the pulmonary artery catheter during hypothermic cardiopulmonary bypass have been documented.en
dc.language.isoenen
dc.titleFailure of continuous cardiac output measurement using the PiCCO Device during induced hypothermia: a case report.en
dc.typeJournal Articleen
dc.identifier.journaltitleCritical Care and Resuscitationen
dc.identifier.affiliationDepartment of Intensive Care Medicine, The Austin Hospital, Heidelberg, Victoria.en
dc.description.pages99-101en
dc.relation.urlhttps://pubmed.ncbi.nlm.nih.gov/16566694en
dc.type.austinJournal Articleen
local.name.researcherBellomo, Rinaldo
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.openairetypeJournal Article-
item.grantfulltextnone-
item.cerifentitytypePublications-
item.fulltextNo Fulltext-
item.languageiso639-1en-
crisitem.author.deptIntensive Care-
crisitem.author.deptData Analytics Research and Evaluation (DARE) Centre-
Appears in Collections:Journal articles
Show simple item record

Page view(s)

6
checked on Oct 3, 2024

Google ScholarTM

Check


Items in AHRO are protected by copyright, with all rights reserved, unless otherwise indicated.